AC Susceptometer/DC Magnetometer
|
Temperature
Range: AC/DC magnetic field Range : from 0.1 to 1600 A/m AC susceptibility sensitivity : 2
* 10 -8 emu 9 * 10 -5 emu
5 Tesla superconducting Magnet Specifications: +/-50000 gauss (+/- 5 Tesla) Field Resolution : < 1 gauss, -50000 gauss to +50000 gauss Remnant field: 15 gauss (<0.5
gauss after demagnetization cycle) |
![]() |
First harmonic ac magnetic susceptibility as a function of temperature, for an MgB2 sample measured at various dc fields |
![]() |
Magnetization
hysteresis loops measured at different temperatures for an MgB2
sample |
![]() |
Magneto-optical
measurement set-up
|
DC
Magnetic field 0 to 1800 Oe Available
magnification: 100x 200x 400x Resolution 1
Oe
|
![]() |
Remnant
state at 5 K after an applied field 150 Oe. |
![]() |
Home-made transport
measurement set-up
| Temperature
range:
Voltage
resolution:
Available characterizations: -
resistence vs temperature in magnetic field up to 6 T |
|
| R vs. T curve. Bias current 10 mA |
![]() |
Microwaves measurement
set-up
click to enlarge
| temperature range: | 3 - 325 K |
| dc magnetic field range: | 0-100 Oe (custom-made high uniformity magnet) 0-6 T (in the Cryogen Free Magnet) |
| main specifications of the ZVK Rohde-Schwarz Vector Network Analyzer | frequency range: 10MHz - 40GHz dynamic range: >80dB measurement time per point: <1.1ms output power (with step attenuators): -90dBm - 0dBm |
apparatus
|
Superconducting
Magnet : |
|